Endotracheal Tube Cuff Inflation Pressure awareness and response to education among anesthesia technicians

Main Article Content

Alparslan Koç
Nurhan Eren

Abstract

Objective: Cuffed endotracheal tubes are used to ensure ventilator compliance and prevent pulmonary aspiration in mechanically ventilated patients. At cuff pressures greater than 40 cmH2O, mucosal irritation, ulceration, hemorrhage, tracheal stenosis, and tracheoesophageal fistula can occur due to increased perfusion pressure of the tracheal mucosa and submucosa. In this study, we compare the awareness of cuff pressure among anesthesia technicians working in the operating room.


Materials and Methods: All anesthesia technicians received a seminar on cuff pressure. An attempt was made to determine the difference between cuff pressures measured before and after the seminar.


Results: A positive correlation was found when the cuff pressure measurement was compared with the first measurement after the training (Cor. Coef.= 0.376). At the first measurement, the mean cuff pressure was 82 cmH20, the lowest pressure was 27, and the highest was 223. At the measurement after completion of the training, the mean pressure was 50, the lowest pressure was 26, and the highest pressure was 105. The difference between the two measurements was statistically significant (p=0.000). Before training, only four technicians (7.40%) inflated below the recommended confidence interval (30 cm H2O), while the remaining 50 technicians (92.6%) inflated above this limit. After training, 11 technicians could inflate below the confidence interval. A statistically significant increase was observed (p < 0.05).


Conclusion: Measuring the cuff pressure of the endotracheal tube was essential to avoid possible complications. Educational seminars on this topic and the provision of cuff meters can avoid these problems.

Downloads

Download data is not yet available.

Article Details

How to Cite
Koç, A., & Eren, N. (2022). Endotracheal Tube Cuff Inflation Pressure awareness and response to education among anesthesia technicians. Medical Science and Discovery, 9(7), 393–396. https://doi.org/10.36472/msd.v9i7.758
Section
Research Article
Received 2022-06-23
Accepted 2022-07-18
Published 2022-07-19

References

Letvin A, Kremer P, Silver PC, Samih N, Reed-Watts P, Kollef MH. Frequent Versus Infrequent Monitoring of Endotracheal Tube Cuff Pressures. 2018;63(5):495-501. DOI: https://doi.org/10.4187/respcare.05926

Nordin U. The trachea and cuff-induced tracheal injury. An experimental study on causative factors and prevention. Acta Otolaryngol Suppl. 1977;345:1-71.

Braz JR, Navarro LH, Takata IH, Nascimento Júnior P. Endotracheal tube cuff pressure: need for precise measurement. Sao Paulo Med J. 1999;117(6):243-7. DOI: https://doi.org/10.1590/S1516-31801999000600004

Sanaie S, Rahmani F, Chokhachian S, Mahmoodpoor A, Rahimi Panahi J, Mehdizadeh Esfanjani R, et al. Comparison of tracheal tube cuff pressure with two technique: fixed volume and minimal leak test techniques. J Cardiovasc Thorac Res. 2019;11(1):48-52. DOI: https://doi.org/10.15171/jcvtr.2019.08

Klainer AS, Turndorf H, Wu WH, Maewal H, Allender P. Surface alterations due to endotracheal intubation. Am J Med. 1975;58(5):674-83. DOI: https://doi.org/10.1016/0002-9343(75)90504-5

Selman Y, Arciniegas R, Sabra JM, Ferreira TD, Arnold DJ. Accuracy of the minimal leak test for endotracheal cuff pressure monitoring. 2020;130(7):1646-50. DOI: https://doi.org/10.1002/lary.28328

Siamdoust SAS, Mohseni M, Memarian AJA, medicine p. Endotracheal tube cuff pressure assessment: Education may improve but not guarantee the safety of palpation technique. 2015;5(3). DOI: https://doi.org/10.5812/aapm.5(3)2015.16163

Pia Hedberg R, Carolina Eklund R, Sandra Högqvist RJAJ. Identification of a very high cuff pressure by manual palpation of the external cuff balloon on an endotracheal tube. 2015;83:179-82.

Sejkorová A, Bolcha M, Beneš J, Kalhous J, Sameš M, Vachata P. Intraoperative Measurement of Endotracheal Tube Cuff Pressure and Its Change During Surgery in Correlation With Recurrent Laryngeal Nerve Palsies, Hoarseness, and Dysphagia After Anterior Cervical Discectomy and Fusion: A Prospective Randomized Controlled Trial.0(0):21925682211046895.

Henderson JJMsa. Airway management in the adult. 2010;7:1573-610. DOI: https://doi.org/10.1016/B978-0-443-06959-8.00050-9

Özcan ATD, Döğer C, But A, Kutlu I, Aksoy ŞM. Comparison of endotracheal tube cuff pressure values before and after training seminar. Journal of Clinical Monitoring and Computing. 2018;32(3):527-31. DOI: https://doi.org/10.1007/s10877-017-0046-7

Seegobin RD, van Hasselt GL. Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. 1984;288(6422):965-8. DOI: https://doi.org/10.1136/bmj.288.6422.965

Inada T, Uesugi F, Kawachi S, Inada K. The tracheal tube with a high-volume, low-pressure cuff at various airway inflation pressures. Eur J Anaesthesiol. 1998;15(6):629-32. DOI: https://doi.org/10.1097/00003643-199811000-00002

Aeppli N, Lindauer B, Steurer MP, Weiss M, Dullenkopf A. Endotracheal tube cuff pressure changes during manual cuff pressure control manoeuvres: An in-vitro assessment. 2019;63(1):55-60. DOI: https://doi.org/10.1111/aas.13249

Coelho RdM, Paiva TTMd, Mathias LAdST. Avaliação in vitro da eficácia de método para limitar a pressão de insuflação dos balonetes das cânulas endotraqueais. Brazilian Journal of Anesthesiology. 2016;66(2):120-5. DOI: https://doi.org/10.1016/j.bjan.2014.06.005

Saraçoğlu A, Dal D, Pehlivan G, Göğüş FY. The Professional Experience of Anaesthesiologists in Proper Inflation of Laryngeal Mask and Endotracheal Tube Cuff. Turk J Anaesthesiol Reanim. 2014;42(5):234-8. DOI: https://doi.org/10.5152/TJAR.2014.87487

Kumar CM, Seet E, Van Zundert TCRV. Measuring endotracheal tube intracuff pressure: no room for complacency. Journal of Clinical Monitoring and Computing. 2021;35(1):3-10. DOI: https://doi.org/10.1007/s10877-020-00501-2

Bulamba F, Kintu A, Ayupo N, Kojjo C, Ssemogerere L, Wabule A, et al. Achieving the Recommended Endotracheal Tube Cuff Pressure: A Randomized Control Study Comparing Loss of Resistance Syringe to Pilot Balloon Palpation. Anesthesiol Res Pract. 2017;2017:2032748. DOI: https://doi.org/10.1155/2017/2032748